Repair and Service
Machine specific safety instructions and hints for repair work
1.2.1
1.2.2
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 3.1 / 2
Agfa Company Confidential

Safety switch

The machine has one safety
switch, located below the top
cover.
It serves as a safety feature if
someone opens the machine e.g.
to remove a film jam.
When the top cover is opened, the
safety switch is activated.
It cuts off the voltage to all motors
and the print head.
For service purposes the safety switch SW1 can be overridden by
means of a service key. In this case there is a risk of injuries by moving
mechanical or current-carrying electrical parts.
Before the printer is closed the service key must be removed again,
otherwise the mechanism of the safety switch may get damaged.

Supplementary ground cable

To meet the regulations for the printer according to the machine standards
EN60601-1 there is usually no Supplementary ground cable required. This
also applies to installations in the patient's room.
Exception:
The printer is located in the patient's room AND is connected to
equipment not corresponding to EN60601-1.
In this case a Supplementary ground
cable must be connected to the
provided 'Equipotential plug'
connector at the back of the printer.
The supplementary ground cable is
not included in the machine
shipment.
Specifications of the cable:
• green/yellow color,
• cross-section of 4 mm
Printed copies of this document are not controlled.
DRYSTAR 5302
(Type 5366/100)
2
.
DD+DIS150.05E
S 8
Figure 1
Equipotential plug
536631JM.cdr
Figure 2
Edition 1, Revision 0
2005-09-09
536631aa.cdr