Repair and Service
Machine specific tools, software tools and auxiliary means
2.5.5.2
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 3.2/34
Agfa Company Confidential
In general: all DICOM host profile
and LPD profile web pages have
an online documentation
included.
It can be viewed by pressing the
HELP button.
Example 2: Define annotations for all inputs
In this example the defaults for all hosts, independent of the AE title are
adapted. Here it is only possible to adapt the annotations.
(1) In the printer homepage select 'Service Engineer tools - Enter service
command session'
(2) Enter username and password
(3) Select 'DICOM profile editor'
(4) Select 'Default site annotations: EDIT'. Refer to Figure 33.