AGFA DRYSTAR 5302 Technical Documentation Manual page 271

Hide thumbs Also See for DRYSTAR 5302:
Table of Contents

Advertisement

DD+DIS150.05E
(14) Leave the "service command session".
(15) Repeat step "2.5.6 Performing the Afssaps / EUREF Calibration Check.
IMPORTANT:
If after three repetitions one or more "Deviation compared to DICOM (%)" values are
still ≥ 10%, start the complete procedure from scratch. Refer to section 2.5.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
05-2009
If this time all "Deviation compared to DICOM (%)" values are < 10%, the
"Afssaps / EUREF calibration procedure" is finished.
If one (or more) "Deviation compared to DICOM (%)" value(s) are ≥ 10%,
repeat steps 2.5.6 and 2.5.7 up to three times. See IMPORTANT note
below.
DOCUMENT CONTROL NOTE:
DRYSTAR 5302
Type 5366/100
Repair and Service
Adjustments and Calibrations
Agfa Company Confidential
Chapter 3.6 / 29

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dd+dis150.05e

Table of Contents