DD+DIS150.05E
HC305010 : thmodule.voltage.low
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
Reason3
Cure3
Part
IMPORTANT:
This fault may have following additional reason and cure:
Reason: Calibration files corrupted.
Cure:
For Software ≥ 2.0.0_C2:
Reset the calibration data for the film calibration of the upper and lower tray, followed
by a film calibration for both trays.
For Software < 2.0.0_C2:
Perform an emergency software installation without preserving the configuration data.
After this set-up the printer as described in chapter 1 of the Service Documentation.
HC305011 : thmodule.thinstall
Type
Context
Action
Reason
Cure
Part
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2011
Failure
W305010
SeeManual
Serial connection to powersupply fails
power down the printer, and power up again
PSU module defect
Replace
MOD09
TH module defect
Replace
MOD05
Defect
05
Replace
TH data is not loaded properly
Replace
MOD05
DOCUMENT CONTROL NOTE:
DRYSTAR 5302
Type 5366/100
Repair and Service
Troubleshooting - Error Codes
Agfa Company Confidential
Chapter 3.3 / 31 of 72