Controls, Connections and Set Up Procedures
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 1 / 24
Agfa Company Confidential
Dmax
Kanamori curve and
Kanamori like value
220
Dmin
0
Interpolation Smooth Factors (Kernel values)
1.1
0.9
5.9
-0.2 -0.7
-1.2
4.9
less sharp
sharp
For details regarding image adjustments refer to the Hardcopy Application
Manual DD+DIS078.05E.
Each of the image adjustment parameters can be set for a specific modality.
Changing a parameter value will change the value of the respective
parameters in the 'DICOM Host profile' for that specific modality.
Printed copies of this document are not controlled.
Drystar 5302
(Typ 5366/100)
Possible Look up table settings
200
100
75
Input
255
Figure 17
0.7
0.1
Cubic Bell
-4.9
Cubic High Res
too sharp
Figure 18
DD+DIS150.05E
Kanamori_like 100 = Kanamori
Kanamori_like 200 = linear
Kanamori_like 220 = logarithmic
536401GM.CDR
Legend
Limits
Default
1.1
0.9
5.9
Reasonable values
536601GL.CDR
Edition 1, Revision 0
0.7
0.1
2005-09-09