DD+DIS150.05E
HC303015 : upperpickupmodule.working
Type
Context
Action
Reason1
Cure1
Part
Reason2
Cure2
HC303016 : upperpickupmodule.passed.high
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC303021 : upperpickupmodule.m7.connection
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2011
Failure
W303015
None
Upper pickup module defect
Replace
MOD03
Film wrinkled in upper tray
Remove wrinkled film
Failure
W303016
None
Upper pickup past sensor jammed
Free upper pickup past sensor
Upper pickup past sensor defect
Replace
MOD03
Failure
W303021
SeeManual
Upper pickup motor not connected
reconnect upper pickup motor
Upper pickup module defect
Replace
MOD03