DD+DIS150.05E
(1)
In the browser screen select menu 'Service Engineer Tools - Enter Service
Verification
Command Session - 3 Print Image - 1 Testimage – upper or lower tray* -
STI2.ZIP': The STI2 test image will be printed.
(2)
From the same menu print the medical_image.zip*
*IMPORTANT: Select the tray with 14" width films.
(3)
Check both images at the light box for dark stripes in the area of the 10" film
edge:
• On the STI2 test image dark stripes may be visible in the area of the 10"
• On the medical test image no dark stripes may be visible in the area of the
The smart profile calibration is performed.
Result
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
05-2009
film edge.
10" film edge.