Machine specific tools, software tools and auxiliary means
What to check in the
infocounter
2.3 Reliability
3.3 Hardware Modification
History
3.4 Software modification
history
5.3 Retries
5.7 Error History
5.8 Error list (sorted by
relative) and
5.9 Error list (sorted by
total):
Printed copies of this document are not controlled.
Drystar 5302
(Type 5366/100)
Infocounters.txt evaluation
Comment
Should normally be not higher than 0.05. Higher
fault rates need a thorough investigation.
By comparing the status of the device with the
available modifications (refer to Technical
Documentation chapter 8) the exact hardware
status can be determined.
By checking the software modification history it
can be determined, whether a recent software
upgrade solved a problem, that occurred quite
often in the error list in the past
Many retries (> 1%, based on relative prints)
have to be investigated: They may reduce the
throughput of the device.
Check which errors occurred in the last days,
how often they appeared as well as the CBF
(cycles between failures) of these errors. This
gives an overview of the current status of the
machine.
Troubleshoot these errors with the help of the
technical documentation, chapter 3.3,
'Troubleshooting'.