DD+DIS150.05E
3.2
Purpose
SW Installation
on empty HD
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Edition 1, Revision 0
2005-09-09
Machine specific tools, software tools and auxiliary means
Caution:
•
Depending on the software patch version it might be possible, that
one of the boards gets re-programmed during the software upgrade.
•
In case this re-programming is interrupted by power loss, this specific
board may become unusable.
•
Do not reboot the printer before requested by the web pages.
Emergency SW Installation on a new, empty Hard Drive
This chapter describes the SW Installation on an empty hard drive, e.g. after
exchange of the hard drive.
(1) Insert the CF card with the printer
"Emergency_Install" software in
the CF card reader
(2) Reboot the printer
(3) At question "Do you want to
preserve config?" select "NO".
Refer to Figure 53.
(4) Follow the further instructions on
the display.
(5) Select menu "5 Restore
Configuration" to reload the
previous configuration when
finished.
(6) Perform a film calibration from
both trays.
Important remark:
After software installation on a new,
empty harddisk, the printer is only in
the same state than before, if an up-
to-date backup is available.
In case the CF card is lost or not
readable the complete printer setup
has to be redone.
After Emergency SW Installation on a
new, empty hard disk without
"Configuration Restore" the printer is
in the same state like it has been
shipped from factory
(default IP address 169.254.10.10)
Printed copies of this document are not controlled.
(Type 5366/100)
Drystar 5302
Repair and Service
초
2
Figure 52
Emergency SW FOUND
ON EXTERNAL CF-card
Install emergency
SW?
YES (V)
NO (X)
Emergency SW Install
Do you want
to preserve config?
YES (V)
NO (X)
536632gc.cdr
Figure 53
Chapter 3.2/51
Agfa Company Confidential
5366_Reg01_017_e.cdr