AGFA DRYSTAR 5302 Technical Documentation Manual page 213

Hide thumbs Also See for DRYSTAR 5302:
Table of Contents

Advertisement

DD+DIS150.05E
HC309003 : powersupplymodule.powerth.current
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
Reason3
Cure3
Part
HC309004 : powersupplymodule.powermotor.current
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC309005 : powersupplymodule.temperature.warning
Type
Context
Action
Reason
Cure
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2011
Defect
E309003
SeeManual
TH power cables short circuited
Check TH power cables
TH module defect
Replace
MOD05
PSU module defect
Replace
MOD09
Defect
E309004
SeeManual
Actuators power cables short circuited
Check actuators power cables
PSU module defect
Replace
MOD09
Failure
E309005
None
PSU temperature high
Cool down the PSU
DOCUMENT CONTROL NOTE:
DRYSTAR 5302
Type 5366/100
Repair and Service
Troubleshooting - Error Codes
Agfa Company Confidential
Chapter 3.3 / 57 of 72