Film Calibration - AGFA DRYSTAR 5302 Technical Documentation Manual

Hide thumbs Also See for DRYSTAR 5302:
Table of Contents

Advertisement

DD+DIS150.05E
2.1

Film Calibration

The film calibration has to be performed in following cases:
When to be
performed
Before Smart TH Profile Calibration. See section 2.4.
During the compliance film calibration. See section 2.5.
After TH replacement
At least each 5000 films (adjustable maintenance action. Refer to chapter 9
Maintenance).
In case of a drift of the maximum density
During the troubleshooting procedure
Determination of the correlation between control of the thermal head and
Purpose
corresponding densities on film
REQUIRED TOOLS:
None
REQUIRED TIME:
10 min.
NOTE:
The film calibration is by default executed automatically at every 5th new film package:
This automatic film calibration can be activated/deactivated in the Key-operator menu
"1. Change settings - 5. Image quality - 1. General - select default for film calibration
(default = ON)". Also the calibration frequency (Default = every 5 film packs) can be
entered there.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
05-2009
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
DRYSTAR 5302
Type 5366/100
Repair and Service
Chapter 3.6 / 7
Agfa Company Confidential