Calibration, Adjustment, Linearization - Sartorius Genius Series Operating Instructions Manual

Electronic analytical and semi-microbalances
Table of Contents

Advertisement

Calibration, Adjustment, Linearization

Purpose
Calibration is the determination of
the difference between the weight
readout and the true weight (mass) of
a sample. Calibration does not entail mak-
ing any changes within the balance.
Adjustment is the correction of this
difference between the measured value dis-
played and the true weight (mass) of
the sample, or the reduction of the
difference to an allowable level within
maximum permissible error limits.
Linearization is the procedure used
to eliminate the deviation from weight
readout plotted on the balance's ideal
characteristic calibration curve and the
actual weight readout. In other words, lin-
earization reduces the amount of error of a
weight displayed to its maximum
permissible error limits. The ideal
characteristic curve is a straight line
plotted between zero load and maximum
load.
Using Verified Balances as Legal
Measuring Instruments in the EU*:
Before using your balance as a legal
measuring instrument, you must perform
"internal calibration" at the place of
installation after the warmup period.
* including the Signatories of the
Agreement on the European Economic
Area
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Features
You can configure whether the
calibration mode
– will be activated according to the
specific setting (external/internal) or
– can be selected by the user after
Cal
pressing the
soft key:
Selection mode
.
Your balance can be calibrated external-
ly: (Balance menu: CAL key function;
Ext. cal./adj.;
menu item
factory-def. wt.
Ext. cal./adj.;
user-defined wt.
Internal
or internally: (
cal./adjustment
).
Adjustment can be performed
– automatically following calibration:
Cal., then auto
adjust.
or
– if desired, the adjustment operation can
be started manually after calibration:
Cal., then manual
adjust
Linearization is performed if you
Internal
have selected
linearization
in the Setup
menu or you have set this using the
Selection mode
.
You can have the balance automatically
display an adjustment prompt after a
certain time interval has elapsed since
the last calibration/adjustment or when
the ambient temperature changes by
a defined amount.
You can also configure the balance to
perform calibration and adjustment
automatically (isoCAL) when the pre-set
time and/or temperature limit is reached
On and reset
application
On
and
without resetting app
(see also page 55).
You can have the calibration/adjustment
results documented in an ISO/GLP-
compliant printout; see page 131.
Factory Settings of the Parameters
Calibration/adjustment mode:
Selection mode
Calibration/adjustment sequence:
Calibrate, then auto
adjust
or
isoCAL function:
resetting application
;
Start automatic adjustment:
Print GLP/GMP calibration/adjustment
Automatic if GLP is
record:
selected
.
On without
isoCAL
47

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents