8.2 Calibration/Adjustment Using an External
Calibration Weight
An external calibration weight is required for this function .
Please note the tolerance of the calibration weight being used .
In verified balances with accuracy class K external calibration/adjustment is
blocked in legal metrology .
Secura
and Quintix
®
can only be accessed when the access switch is open .
Make sure that the weighing pan is empty .
t
Select
A
t
Select the menu key at the bottom left of the display in the current application .
t
y The menu appears .
Mixing
Select the CAL button to access the calibration function .
t
Calibrate balance
y The
Select
CAL-Extern
t
y Practum
®
Select the value of your calibration weight .
t
y A message appears on the display, prompting the user to place the calibration
weight on the weighing pan .
Place the calibration weight on the weighing pan .
t
y Calibration starts automatically .
y A report appears on the display (Secura
11:03
The report indicates the deviations identified during calibration . The report also
gives the result of the adjustment process .
Close the report window: Select
t
y The balance is now calibrated and adjusted .
a
Information on potential errors can be found in the "Status Messages" chapter on
page 91 .
In verified balances with accuracy class k, external adjustment
®
to zero the balance .
window appears .
.
models: Calibration values appear on the display .
<
.
User Manual Secura
Calibration and Adjustment
®
and Quintix
®
only) .
®
, Quintix
®
, Practum
75
®