Increasing- Load Test (Phase 2) - Datalogic MAGELLAN 8200 Product Reference Manual

Table of Contents

Advertisement

Increasing- Load Test (Phase 2)

The upper limit of the scale is configurable according to POS interface type
and may not necessarily be set at 15 kilograms, which is the standard set-
ting. For this test, continue to place weights in 2.50 kilogram increments
only up to the upper weight limit set for your scale.
NOTE
Product Reference Guide
Figure 5-3. European Shift Test (Metric)
After completing the Shift Test, you must complete the Increasing Load
Test using 10.00, 12.50 and 15 kilograms of weight.
1. Place 10.00 kilograms in the center of the weighing surface and
check that the display reads between 9.990 kg and 10.010 kg.
2. Place an additional 2.50 kilograms in the center of the weighing
surface and check that the display reads between 12.490 kg and
12.510 kg.
A
B
CENTER
C
D
5-15