K-TEK KTEK9000 Instructions For Use Manual page 7

Curve spine frame

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
INSTRUCTIONS FOR USE
4. Ensure the leafs are captured under the lower side edge of the frame.
3.4 Positioning the Patient:
Position the patient onto the operating room table according to the requirements of the
surgeon and facility protocol.
1. Connect the Crank Handle. Turn Counter-clockwise to curve the pads. Turn Clock-
wise to flatten the pads.
3.5 Device Maintenance:
Make sure that all labels are installed and can be read. Replace labels as necessary by
using a plastic scraper to remove the label. Use an alcohol wipe to remove any adhe-
sive residue.
Contact Kyra Medical, Inc., if you need to repair or replace the device contact us using
the information from the contact details section.
3.6 Cleaning and Disinfection:
WARNING:
• After each use, clean the device as directed in this instruction for use.
• Do not submerge the device in liquid. Equipment damage can occur.
• Use caution in areas where liquid can get into the mechanism.
• Do not use bleach or products that contain bleach to clean the device. Injury or
equipment damage can occur.
• Make sure that the device is dry before you store it or use it again.
Wipes:
• Wipes may contain up to 2% sodium hypochlorite.
• Wipes may contain benzalkonium chloride (< or = 0.6% conc.) and didecyl dimethyl
ammonium chloride (up to 0.6% conc.)
©2023 Kyra Medical, Inc.