10 About Calibration
C
T
ALIBRATE
EMPERATURE
This operation typically is performed by a Micromeritics Service Representative.
Press Alt + ., then press CHOICE until Temperature is displayed.
Display Text
Calibrate Temperature
Temperature:
Z
P
ERO THE
RESSURE
The pressure transducer zeroes automatically before each cycle in an analysis or calibration;
therefore, it is not necessary to zero the unit for these operations. Typically, this function is not
required unless manually performing analyses for an extended period of time.
Press Alt + 0 to access the Zero function.
Display Text
Zero Type?
Pressure
Zero Volume?
Chamber insert?
None
Does not display for
Pressure
Zero Volume
[ENTER] to start
[ESCAPE] to cancel
or
Zero Pressure
[ENTER] to start
[ESCAPE] to cancel
10 - 20
Description
Enter the temperature obtained from a reference temperature sensor.
Press ENTER to automatically return to the Reload prompt. Press ALT
+ CLEAR to cancel the operation.
T
& C
RANSDUCER
Description
Allows zeroing of the pressure transducer or the volume offset. Press
CHOICE until the option is displayed, then press ENTER. If choosing
Pressure, the next prompt is not displayed. Choices are Pressure and
Volume. Ensure the sample chamber is empty if selecting Volume.
Displays when Volume is selected. Select Yes at the Request insert
prompt in Report Options. Press CHOICE until the appropriate insert is
displayed, then press ENTER.
3
None, 0.1 cm
(for 1 cm
n
3
None, 3.5 cm
, 1.0 cm
n
3
None, 35 cm
, 10.0 cm
n
3
None, 650 cm
, 1300 cm
n
Press ENTER to begin the zero operation or Alt + CLEAR to cancel
the operation.
If zeroing the Volume, the purges and cycles specified in the current
analysis parameters are used to calculate the volume offset.
After the operation is complete, the pycnometer returns to the Reload
prompt.
V
HAMBER
OLUME
3
unit)
3
3
(for 10 cm
unit)
3
3
(for 100 cm
unit)
3
3
(for 2000 cm
pycnometer)
AccuPyc II 1345 Operator Manual
134-42851-01 (Rev A) — Jan 2021