Assembly and preparation
Connecting the gas supply
WARNING
Risk due to gas supply failure
All gas supplies (central gas supply, gas
cylinders) must be correctly connected since
otherwise the backup system (gas cylinders)
will not be available if gas supply fails.
– Make sure that all compressed gas hoses
are correctly connected to the rear side of
the device.
– After connecting the gas supply, check for
correct function.
– Even when the anesthesia machine is
connected to the central gas supply, the
gas cylinders should remain at the device
with valves closed as backup.
WARNING
Risk of contamination of ambient air and risk
of fire/explosion
O
or N
O can get into the ambient air as a
2
2
result of leakages.
– Make sure the compressed gas hoses are
connected properly.
– Avoid and remedy all leakages.
– Ensure sufficient ventilation of the room.
– Use an anesthetic gas scavenging
system.
56
Central gas supply
Version for 3 gases
WARNING
Risk to patient through incorrect gas
connection
If the connections of the central gas supply
hoses between the central gas supply and
Fabius are interchanged, serious accidents
can occur.
If Fabius is connected to the central gas
supply, it must be checked whether the marks
on the hoses agree with the connections of
the central gas supply.
A
2.8-6.0 kPa
x 100 (bar)
N2O
2.8-6.0 kPa
x 100 (bar)
AIR
2.8-6.0 kPa
x 100 (bar)
O2
1 Screw the connection piece of every individual
central gas supply hose hand-tight on the
corresponding connection (A) on the device
side.
2 Plug the probes of the central gas supply
hoses in the appropriate wall terminal units (B).
3 Check if all central gas supply hoses are
correctly connected.
Instructions for use Fabius GS premium SW 3.n
B
2.8-6.0 kPa
x 100 (bar)
N2O
2.8-6.0 kPa
x 100 (bar)
AIR
2.8-6.0 kPa
x 100 (bar)
O2