Connecting Gas Probes For The Gas Supply (Applies Only To Dräger Din Terminal Units) - Dräger Gemina DUO Instructions For Use Manual

Wall supply unit
Table of Contents

Advertisement

Connecting gas probes for the gas supply (applies only to
Dräger DIN terminal units)
WARNING
Increased fire risk
Terminal units must not come into contact
with oil, grease or flammable liquids.
CAUTION
Risk of device malfunction
Also observe the respective manufacturer's
instructions when using the gas terminal units.
A
B
Park position
A Push the probe into the gas terminal unit until it
engages for the first time (park position).
Operating position
B Push the probe in until it engages for the
second time.
Instructions for use GeminaDUO
A
B
Interrupting the gas supply
A Apply slight pressure to release bushing, probe
will return to parking position.
Fully disconnecting the plug
B Push the release bushing with more force and
at the same time remove the probe from the
terminal unit.