A-1. Calibration Mode Menu Structure 1/2

Zero
Calibration
20 seconds countdown
with zero calibration icon blinking
Zero Cal.
Success (all)
< ON/OFF > for 5 seconds
while "
" is blinking for 5 seconds
Span Gas
Information
wait for 10 seconds
Span Gas
Search
span gas search
for up to 30 seconds
Span
Calibration
60 seconds countdown
with span calibration icon blinking
X4 Series
Zero Cal.
Fail
< ON/OFF >
within 10 seconds to
activate the function
for setting span gas
concentration
< ON/OFF >
to save H
S value
2
< 3 Gas Calibration >
CH
4
CO
H
S
2
balance
O
/N
2
2
No gas
found
Zero Cal.
Fail (some)
Set Span Gas
Concentration
No Gas
Found
Zero Cal.
Fail (all)
< UP >
to increase
or
< DOWN >
to decrease
the value
---------------------
< ON/OFF >
to save the value
and
scroll to the next gas
(Exp / CO / H
S)
2