Power Supply Assembly Replacement

Tools Required

Torque screwdriver, TORX T20
Torque screwdriver, TORX T8
Torque screwdriver, TORX T10
Box torque wrench, 5/8 inch
Box torque wrench, 5.5 mm

Removal Procedure

Refer to Figure 6-9 for this procedure.
Step 1. Remove the 3/4 rack module top as described in "3/4 Rack Module Top Replacement" on
page 90.
Step 2. Remove the PPMC PCA as described in "PPMC PCA Replacement" on page 101.
Step 3. Remove the A1 Mother Board as described in "A1 Mother Board Replacement" on
page 103.
Step 4. Disconnect the cable of the inlet (item 1) from the Power Supply Assembly.
Step 5. Remove the four 5.5 mm Standoff (item 2).
Step 6. Remove the TORX T10 screw (item 3) fastening the CPU cover.
Step 7. Remove the CPU cover (item 4).
Step 8. Remove the Power Supply Assembly with the cable power control.
Step 9. Disconnect the cable power control (item 5) from the Power Supply Assembly.
NOTE
Do not pull out the power cable (P/N : E4980-61601) from the Power Supply Assembly
when you detach the Power Supply Assembly. There is a possibility that the Power Supply
Assembly could break when the power cable is forcibly pulled out. (Refer to sub-figure in
Figure 6-9.)

Replacement Procedure

Step 1. Reverse the order of the removal procedure.
NOTE
When you fasten the CPU cover with the TORX T10 screw, set screwdriver's torque to 8.7
in-lb.
When you fasten the Power Supply Assembly with the 5.5 mm Standoff, set box torque
wrench's torque to 7 in-lb.
When you replace the Power Supply Assembly with a new one, use a new power cable
(P/N : E4980-61601).
Chapter 6
Replacement Procedure
Power Supply Assembly Replacement
105