Step 2. Remove the Front Panel Assembly
Refer to
Figure 3
for this procedure.
1. With a T-10 TORX driver, remove the six screws (item
Before removing the front panel from the analyzer, lift and support the front of
CAUTION
the analyzer chassis.
2. Slide the front panel over the test port connectors.
3. Disconnect the front panel interface ribbon cable (item
from the analyzer.
Figure 3
Front Panel Assembly Removal
Installation Note E8801-90024
) from the sides of the frame.
). The front panel is now free
9