Electrical connection
28
Signal damping
Max. 9 dB over the entire length of the cable cross-section
Shield
Copper braided shielding or braided shielding with foil shield. When grounding
the cable shield, observe the grounding concept of the plant.
For further information on planning and installing PROFIBUS networks see:
Operating Instructions "PROFIBUS DP/PA: Guidelines for planning and commissioning"
(BA00034S)
Cable diameter
• Cable glands supplied:
M20 × 1.5 with cable ⌀ 6 to 12 mm (0.24 to 0.47 in)
• Spring terminals:
Wire cross-sections 0.5 to 2.5 mm
7.2.3
Terminal assignment
Transmitter
PROFIBUS DP connection version
For use in the non-hazardous area and Zone 2/Div. 2
Order code for "Output", option L
Depending on the housing version, the transmitters can be ordered with terminals or
device plugs.
Connection methods available
Order code
"Housing"
Output
Options
Terminals
A, B
Options
Device plug
A, B
connectors
→ 29
Options
Device plug
A, B, C
connectors
→ 29
Order code for "Housing":
• Option A: compact, coated aluminum
• Option B: compact, hygienic, stainless
• Option C ultra-compact, hygienic, stainless
Proline Cubemass C 100 PROFIBUS DP
2
(20 to 14 AWG)
Possible options for order code
Power
"Electrical connection"
supply
Terminals
• Option A: coupling M20x1
• Option B: thread M20x1
• Option C: thread G ½"
• Option D: thread NPT ½"
Terminals
• Option L: plug M12x1 + thread NPT ½"
• Option N: plug M12x1 + coupling M20
• Option P: plug M12x1 + thread G ½"
• Option U: plug M12x1 + thread M20
Device plug
Option Q: 2 x plug M12x1
connectors
→ 29
Endress+Hauser
Need help?
Do you have a question about the Proline Cubemass C 100 PROFIBUS DP and is the answer not in the manual?