Commissioning
Parameter
Corrected volume flow calculation
External reference density
Fixed reference density
Reference temperature
72
Navigation
"Setup" menu → Advanced setup → Calculated values
Calculated values
Corrected volume flow calculation
Parameter overview with brief description
Prerequisite
In the Corrected volume flow
calculation parameter, the
External reference density
option is selected.
The Fixed reference density
option is selected in the
Corrected volume flow
calculation parameter
parameter.
The Calculated reference
density option is selected in
the Corrected volume flow
calculation parameter
parameter.
Corrected volume flow calculation
External reference density
Fixed reference density
Reference temperature
Linear expansion coefficient
Square expansion coefficient
Description
Select reference density for
calculating the corrected
volume flow.
Shows external reference
density.
Enter fixed value for reference
density.
Enter reference temperature
for calculating the reference
density.
Proline Cubemass C 100 PROFIBUS DP
Selection / User
Factory setting
interface / User
entry
• Fixed reference
Calculated reference
density
density
• Calculated
reference density
• Reference density
by API table 53
• External reference
density
Floating point
number with sign
Positive floating-
1 kg/Nl
point number
–273.15 to 99 999 °C Country-specific:
• +20 °C
• +68 °F
Endress+Hauser