Endress+Hauser Stamolys CA71CU Operating Instructions Manual

Endress+Hauser Stamolys CA71CU Operating Instructions Manual

Analyzer for photometric determination of copper
Hide thumbs Also See for Stamolys CA71CU:

Advertisement

Quick Links

Operating Instructions
Stamolys CA71CU
Analyzer for Photometric Determination of Copper
BA359C/07/en/10.06
51512322
Valid as of:
Software version 5.9

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Stamolys CA71CU and is the answer not in the manual?

Questions and answers

Summary of Contents for Endress+Hauser Stamolys CA71CU

  • Page 1 Operating Instructions Stamolys CA71CU Analyzer for Photometric Determination of Copper BA359C/07/en/10.06 51512322 Valid as of: Software version 5.9...
  • Page 2 → ä 36, → ä 60 Maintenance schedule → ä 36 Replacing consumables and wear parts → ä 42, → ä 56 Accessories Æ → ä 43 Troubleshooting → ä 45, → ä 56 Spare parts Æ → ä 49 Technical data Endress+Hauser...
  • Page 3: Table Of Contents

    Stamolys CA71CU Table of contents Safety instructions ....4 Dilution module ......42 Additional accessories .
  • Page 4: Safety Instructions

    Safety instructions Stamolys CA71CU Safety instructions Designated use The analyzer is a compact photometric analytical system. It is designed for monitoring copper content in industrial applications. CA71 is particularly suitable for the following applications: • Monitoring copper in industrial sewage treatment plants •...
  • Page 5: Notes On Safety Icons And Icons

    Stamolys CA71CU Safety instructions Notes on safety icons and icons Warning! This symbol alerts you to hazards. They could cause serious injuries as well as damage to the instrument if ignored. Caution! " This symbol alerts you to possible faults which could arise from incorrect operation. They could cause damage to the instrument if ignored.
  • Page 6: Identification

    Identification Stamolys CA71CU Identification Device designation 2.1.1 Nameplate Compare the order code on the nameplate (on the analyzer) to the product structure and your order. You can read the following information from the nameplate: • Order code (device version) • Serial number •...
  • Page 7: Scope Of Delivery

    Stamolys CA71CU Identification Scope of delivery Note! Reagents must be ordered separately for version CA71XX-XXXXXX1. Inactive reagents form part of the scope of delivery for all other versions. They must be mixed before use. Please refer to the guidelines enclosed for this purpose.
  • Page 8: Installation

    Installation Stamolys CA71CU Installation The analyzer at a glance a0006574 Fig. 2: Analyzer (cabinet version, without hoses) Reagent pump (P2), supply from canister Valve V2 Display Outlet for sample or reagent mixture Serial interface RS 232 Valve V1 Photometer cell...
  • Page 9: Incoming Acceptance, Transport, Storage

    Stamolys CA71CU Installation Incoming acceptance, transport, storage • Make sure the packaging is undamaged! Inform the supplier about damage to the packaging. Keep the damaged packaging until the matter has been settled. • Make sure the contents are undamaged! Inform the supplier about damage to the delivery contents.
  • Page 10 Installation Stamolys CA71CU Version with GFR housing 163 (6.42) 250 (9.84) 160 (6.30) 140 (5.51) 100(3.94) 330 (13.0) 60 (2.36) 7-11/ 0.28 - 0.43 (M 6) 200 (7.87) Ø 26 (1.02) 326 (12.8) Ø 16 (0.63) 360 (14.2) 436 (17.2) Ø...
  • Page 11 Stamolys CA71CU Installation Note! For the reagents, you need an additional base at max. 35 cm (13.8 inch) under the pumps in the open version. The dimensions of the reagent bottles are as follows: 90 x 90 x 215 mm (3.54 x 3.54 x 8.46 inch).
  • Page 12: Installation Instructions

    Installation Stamolys CA71CU Adjusting the level monitoring You have to adjust conductive level monitoring depending on whether you want to connect one, two or three analyzers to this single collecting vessel. Depending on the application, mount the right adjusting pin (→ å 8, → å 9, item 2) or do not use any adjusting pin.
  • Page 13 Stamolys CA71CU Installation a0006362 Fig. 11: Valves and valve hoses V1-2 Valves Outlet hose Channel switchover Hose, valve 2 at front, standard To sample pump Hose, valve 2 at rear, cleaner Connecting hose to valve 1, at rear Hose, valve 1 at front, sample Secure the hose boxes in the appropriate pump holders (→...
  • Page 14: Installation Examples

    Installation Stamolys CA71CU Installation examples 3.5.1 CAT 430 or ultrafiltration provided by customer and two CA 71 units • Permeate can contain air bubbles (CAT430) or is free of bubbles (ultrafiltration provided by customer) • Distance between the analyzers as short as...
  • Page 15: Wiring

    Stamolys CA71CU Wiring Wiring Electrical connection Warning! • The electrical connection may only be established by an electrical technician. • The electrical technician must have read and understood these Operating Instructions and must follow the instructions they contain. • Before beginning the connection work, ensure that no voltage is applied to the cables.
  • Page 16 Wiring Stamolys CA71CU 4.1.2 Terminal assignment " Caution! The following diagram shows an example of the connection compartment sticker (→ å 16). The terminal assignment and cable colors can deviate from the actual assignment and colors! Only use the terminal assignment of the stick-on label in the device (→ å 17) to connect your...
  • Page 17 Stamolys CA71CU Wiring Function Designation Terminal Terminal One-channel version Two-channel version Mains Alarm value 1, channel 1 Alarm value 2, channel 1 – Alarm value 1, channel 2 – – – Alarm value 2, channel 2 – – Fault Unassigned Analog output 1 –...
  • Page 18: Signal Connections

    Wiring Stamolys CA71CU Signal connections 4.2.1 Shielding of the analog outputs The interference suppressor dampens interference from electromagnetic effects on control, power and signal lines. After connecting the data transfer cables, clip the interference suppressor (in scope of delivery) onto the cable cores (not onto the outer insulation of the cable!).
  • Page 19: Switching Contacts

    Stamolys CA71CU Wiring 4.2.3 Two-channel version Connection Designation Function Leak Liquid has collected in the drip pan Signal inputs No sample No sample available, measurement is not started, display flashes AV 1 - 1 Alarm value 1, channel 1 exceeded or undershot...
  • Page 20: Serial Interface

    Wiring Stamolys CA71CU Terminal connection for Terminal connection for Terminal connection for condition fulfilled condition not fulfilled power off Channel 1/2 measurement end A = NO configured R = NC configured Note! Condition met means: • AV 1: concentration > alarm value 1 •...
  • Page 21: Post-Connection Check

    Stamolys CA71CU Wiring Post-connection check Carry out the following checks after electrical connection: Device status and specifications Note Are the analyzer or cables damaged externally? Visual inspection Electrical connection Note Does the supply voltage correspond to the data on the nameplate?
  • Page 22: Operation

    Operation Stamolys CA71CU Operation Display and operating elements M CE RS 232 a0001679 Fig. 19: Display and operating elements LED display (measured value) LC display (measured value + status) Serial interface RS 232 Operating keys with indicator LEDs Local operation...
  • Page 23 Stamolys CA71CU Operation 5.2.1 Main menu Access the main menu by holding down the key until AUTO MEASURING is displayed. The main menu items and information on them are explained in the following table. Menu item Display Info AUTO MEASURING...
  • Page 24 Operation Stamolys CA71CU 5.2.3 CONFIGURATION Note! Some settings that can be made in this menu affect the values in the PARAMETER ENTRY menu. In view of this, run through the CONFIGURATION menu first during initial commissioning. Menu item Range of adjustment...
  • Page 25 Stamolys CA71CU Operation Menu item Range of adjustment Display Info (factory settings in bold) Selection for the measuring range scope for channel 1. If the concentration measuring range is 0 to x mg/l, either Analog output 1 4 mA or 0 mA corresponds to 0 mg/l. The end of the measuring range is the same in both cases at 20 mA.
  • Page 26 Operation Stamolys CA71CU 5.2.4 PARAMETER ENTRY Menu item Range of adjustment Display Info (factory settings in bold) Measuring range The specified concentration is allocated a value of 0 or 4 start 1 mA at analog output 1 CU-A: 0.1 to 2 mg/l / 0.00 mg/l CU-B: 0.2 to 5 mg/l / 0.00 mg/l...
  • Page 27 Stamolys CA71CU Operation Menu item Range of adjustment Display Info (factory settings in bold) Time of 1st flushing (DD.MM.YY, time hh.mm). After each change the instrument does not wait for the flushing interval. If the flushing is to start immediately, set the time in the past.
  • Page 28 Operation Stamolys CA71CU 5.2.7 SERVICE Menu item Display Info "Virtual switching board" Various valve and pump combinations can be selected. The setting options are: – Valve 1: P (sample) or S (standard) – Valve 2: S (standard) or R (cleaner) –...
  • Page 29: Calibration

    Stamolys CA71CU Operation 5.2.9 DATA MEMORY-calibration data Note! You get to the data memory calibration data by means of the CONFIGURATION menu. In the "Calibration factor" menu item, press the key. Menu item Display Info This data memory contains the last 100 calibration factors with date and time.
  • Page 30 Operation Stamolys CA71CU 5.3.2 Calibration example You want to activate an immediate calibration (e.g. after you have replaced the reagents). You have replaced the reagents and filled the hoses again (no air bubbles). The analyzer is in the measuring mode.
  • Page 31: Commissioning

    Stamolys CA71CU Commissioning Commissioning Function check Warning! • Check that all connections have been made correctly. Check, in particular, that all hose connections are secure so that no leaks occur. • Ensure that the line voltage corresponds to the voltage specified on the nameplate.
  • Page 32 Commissioning Stamolys CA71CU Function Duration [s] Range of adjustment Calibration Flushing (standard) 3 x 15 Delay to standard 20 to 999 CONFIGURATION / "Delay to sample" Stabilization 1. measurement (basis) SERVICE / "Pumps and valves" Fill mixture Reaction see Technical...
  • Page 33 Stamolys CA71CU Commissioning 6.2.3 Using a dilution module You can extend the measuring range of your analyzer with the sample dilution module. How it works: • The analyzer determines the absorption of the sample at the specified wavelength. It does not matter whether the sample was diluted previously or not.
  • Page 34 Commissioning Stamolys CA71CU a0006622 Fig. 20: Dilution module Hose box of the original sample pump Hose box for additional sample (standard) Hose box for dilution water Sample supply to photometer Overflow (in outlet of analyzer) Sample hose (sample or standard) Dilution water inlet (to be provided by the client) T-section 3.2 x 3.2 x 3.2 mm (0.13 x 0.13 x 0.13 inch)
  • Page 35 Stamolys CA71CU Commissioning PARAMETER ENTRY Measuring range start 1 / measuring range start 2 Enter values that refer to the concentration of the undiluted sample. Measuring range end 1 / If you are using a dilution module with the factor 10, for example, multiply the settings to measuring range end 2 date (without the dilution module) by 10.
  • Page 36: Maintenance

    Maintenance Stamolys CA71CU Maintenance " Caution! You may not carry out any procedures not listed in the following chapters yourself. This work may only be carried out by Service. Maintenance schedule All the maintenance tasks that have to be carried out during normal operation of the analyzer are explained below.
  • Page 37: Replacing Pump Hoses

    Stamolys CA71CU Maintenance 7.2.1 Checking reagents Check the concentration of the standard in the laboratory. Adapt the values (PARAMETER ENTRY, calibration solution") or replace the standard solution. Mix 10 ml standard solution and 5 ml of each reagent in a beaker. You must replace the reagents if there is no visible coloration after max.
  • Page 38 Maintenance Stamolys CA71CU Roller head Upper retainer of hose box Lower retainer of hose box Guide at pump hose Nipple with guide a0001652 Fig. 21: Reagent pump Installing the new hoses Fit the new hose on the hose box. First pull the hose downwards at both ends and then push the guide on the hose into that of the nipple on the hose box.
  • Page 39: Replacing Valve Hoses

    Stamolys CA71CU Maintenance Tighten the screw another complete rotation. a0001651 Fig. 23: Sample pump Adjuster screw for the contact pressure " Caution! Adjust the contact pressure of the reagent hoses so that no sample is pumped into the reagent. Otherwise the reagent becomes immediately unusable.
  • Page 40: Replacing The Static Mixer

    Maintenance Stamolys CA71CU Replacing the static mixer Proceed as follows to replace the mixer (see "Troubleshooting/spare parts"): Rinse first with water and then with air (see SERVICE). Unscrew the four screws on the photometer housing and remove it. Disconnect the mixer from the photometer and from the T-section under the photometer housing or release the mixer from the retainer.
  • Page 41: Cleaning

    Stamolys CA71CU Maintenance Cleaning " Caution! When cleaning, ensure that you do not damage the nameplate on the analyzer. Do not use any solvent-based cleaning agents. To clean the analyzer housing, proceed as follows: • Stainless steel housing (stainless steel SS 1.4301 (AISI 304)): With a lint-free cloth and Glittol RG 10.51...
  • Page 42: Accessories

    Accessories Stamolys CA71CU Accessories Note! The following section provides you with information on the accessories which could be supplied at the time this documentation was issued. Please contact your local Service Center for accessories that are not listed here. Collecting vessel –...
  • Page 43: Troubleshooting

    Stamolys CA71CU Troubleshooting Troubleshooting Troubleshooting instructions Although the analyzer is not very prone to faults due to its simple design, problems at the measuring point can, of course, not be completely ruled out. Possible errors, their causes and their possible remedies are thus listed below.
  • Page 44 Troubleshooting Stamolys CA71CU 9.1.2 System errors/process errors without messages Error Possible cause Tests and / or corrective measures Reagents contaminated Simple check: Mix 5-10 ml of standard solution and 5 ml of reagent in a beaker. If it does not change color or aged after max.
  • Page 45: Spare Parts

    Stamolys CA71CU Troubleshooting Error Possible cause Tests and / or corrective measures Time of 1st The date must be between 01.01.1996 and the current date. measurement not "Measurement" display reached flashes Interval not expired Change parameter setting. Time of 1st calibration The date must be between 01.01.1996 and the current date.
  • Page 46: Software History

    Troubleshooting Stamolys CA71CU 9.2.2 Spare parts for sample and reagent transportation Item Spare part Order number Norprene hose, 1.6 mm 51504116 C-Flex hose, 3.2 mm (permeate inlet and overflow with sample collector.) 51504114 C-Flex hose ID 6.4 mm 51504115 C-Flex hose, 1.5 mm 51512535 Hose connector 1.6 mm x 1.6 mm (10 pieces)
  • Page 47 Stamolys CA71CU Troubleshooting Date Version Changes in the software Documentation 09/2005 Extension BA356C/07/xx/07.06 • New parameter PH-E • Altered factory settings for SI-A and AM-B • Characteristics for FE-B, CU-A and CR-A extended • Measuring range limitation increased to 50000 for all parameters and engineering units •...
  • Page 48: Return

    Troubleshooting Stamolys CA71CU Return If the analyzer requires repair, please send it cleaned to your sales center. Please use the original packaging when returning the device. Please enclose the completed "Declaration of Contamination and Cleaning" form (copy the second last page of these Operating Instructions) with the packaging and also the shipping documents. The...
  • Page 49: Technical Data

    Stamolys CA71CU Technical data Technical data 10.1 Input Measured variable Cu (II) [mg/l] Measuring range • CU-A 0.10 to 2.00 mg/l • CU-B 0.20 to 5.00 mg/l Wave length 565 nm Reference wave length 880 nm 10.2 Output Output signal...
  • Page 50: Environment

    Technical data Stamolys CA71CU Maintenance interval 6 months (typically) Servicing requirements 15 minutes / week (typically) 10.5 Environment 5 to 40 °C (40 to 100 °F), severe fluctuations must be avoided Ambient temperature range Humidity Below the condensation limit, installation in usual, clean rooms...
  • Page 51 Stamolys CA71CU Technical data Endress+Hauser...
  • Page 52: Appendix

    Appendix Stamolys CA71CU Appendix 11.1 Operating matrix One-channel version a0001907-en Endress+Hauser...
  • Page 53 Stamolys CA71CU Appendix a0001908-en Endress+Hauser...
  • Page 54 Appendix Stamolys CA71CU Two-channel version a0001909-en Endress+Hauser...
  • Page 55 Stamolys CA71CU Appendix a0001910-en Endress+Hauser...
  • Page 56: Ordering Forms

    Appendix Stamolys CA71CU 11.2 Ordering forms 11.2.1 Reagents and accessories To Fax No: Fax for ordering reagents To (address of your sales center) From (billing address) Company: Street: Zip code / Town: Fax / Telephone: Address for delivery (if not the above address)
  • Page 57 Stamolys CA71CU Appendix 11.2.2 Ordering wear parts To Fax No: Fax for ordering wear parts To (address of your sales center) From (billing address) Company: Street: Zip code / Town: Fax / Telephone: Address for delivery (if not the above address)
  • Page 58: Analyzer Settings

    Appendix Stamolys CA71CU 11.3 Analyzer settings Place: Type: Analyzer serial No.: Photometer serial No.: Software version: Date: Photometer type: Measuring unit: Calibration factor: C-offset: • mg/l • μg/l Dilution: Delay to sample: Analog output: • 0-20 mA • 4-20 mA AV 1: •...
  • Page 59 Stamolys CA71CU Appendix Submenu Error mask: MB >: MBE: Post rinsing: Filling time: Reaction time: RPM: K floating mean: Points mg/l / μg/l F 1: mg/l / μg/l F 2: mg/l / μg/l F 3: mg/l / μg/l F 4: mg/l / μg/l...
  • Page 60: Maintenance Schedule

    Appendix Stamolys CA71CU 11.4 Maintenance schedule a0001911-en Endress+Hauser...
  • Page 61: Index

    Stamolys CA71CU Index Incoming acceptance ......9 Input ......... 49 Accessories .
  • Page 62 Stamolys CA71CU Quality certificate ....... . 7 Reaching the terminal block......15 Reagents .
  • Page 64 www.endress.com/worldwide BA359C/07/en/10.06 Printed in Germany / FM+SGML 6.0 / DT...