Technical data
Installation conditions
Ambient temperature
range
Storage temperature
Climate class
Degree of protection
Vibration resistance
Shock resistance
Shock resistance
Interior cleaning
116
16.7
Installation
→ 19
16.8
Environment
→ 21→ 21
Temperature tables
Observe the interdependencies between the permitted ambient and fluid
temperatures when operating the device in hazardous areas.
For detailed information on the temperature tables, see the separate document
entitled "Safety Instructions" (XA) for the device.
–40 to +80 °C (–40 to +176 °F), preferably at +20 °C (+68 °F) (standard version)
–50 to +80 °C (–58 to +176 °F) (Order code for "Test, certificate", option JM)
DIN EN 60068-2-38 (test Z/AD)
Transmitter and sensor
• As standard: IP66/67, type 4X enclosure
• When housing is open: IP20, type 1 enclosure
• Display module: IP20, type 1 enclosure
• Oscillation, sinusoidal, following IEC 60068-2-6
• 2 to 8.4 Hz, 3.5 mm peak
• 8.4 to 2 000 Hz, 1 g peak
• Oscillation, broadband noise following IEC 60068-2-64
• 10 to 200 Hz, 0.003 g
• 200 to 2 000 Hz, 0.001 g
• Total: 1.54 g rms
Shock, half-sine according to IEC 60068-2-27
6 ms 30 g
Shock due to rough handling following IEC 60068-2-31
• SIP cleaning
• CIP cleaning
2
/Hz
2
/Hz