Capacitance: core/shield
Maximum 1 000 nF for Zone 1, Class I, Division 1
L/R
Maximum 24 µH/Ω for Zone 1, Class I, Division 1
7.2.3

Terminal assignment

Transmitter: supply voltage, input/outputs
The terminal assignment of the inputs and outputs depends on the individual order
version of the device. The device-specific terminal assignment is documented on an
adhesive label in the terminal cover.
Supply voltage
1 (+)
2 (–)
26 (+)
Terminal assignment of the remote display and operating module →  39.
7.2.4

Preparing the measuring device

NOTICE
Insufficient sealing of the housing!
Operational reliability of the measuring device could be compromised.
Use suitable cable glands corresponding to the degree of protection.
1. Remove dummy plug if present.
2. If the measuring device is supplied without cable glands:
Provide suitable cable gland for corresponding connection cable.
3. If the measuring device is supplied with cable glands:
Observe the requirements for connection cables.
7.3

Connecting the measuring device

NOTICE
An incorrect connection compromises electrical safety!
Only properly trained specialist staff may perform electrical connection work.
Observe applicable federal/national installation codes and regulations.
Comply with local workplace safety regulations.
Always connect the protective ground cable  before connecting additional cables.
When using in potentially explosive atmospheres, observe the information in the
device-specific Ex documentation.
Input/output
Input/output
1
27 (–)
24 (+)
Device-specific terminal assignment: adhesive label in terminal cover.
Electrical connection
Input/output
2
3
25 (–)
22 (+)
23 (–)