Establishing A Connection - Endress+Hauser Proline Cubemass C 100 PROFIBUS DP Operating Instructions Manual

Coriolis flowmeter
Hide thumbs Also See for Proline Cubemass C 100 PROFIBUS DP:
Table of Contents

Advertisement

Operation options
42
8.4.3

Establishing a connection

Via service interface (CDI-RJ45)
Preparing the measuring device
Configuring the Internet protocol of the computer
The following information refers to the default Ethernet settings of the device.
IP address of the device: 192.168.1.212 (factory setting)
1. Switch on the measuring device.
2. Connect to the computer using a cable →  122.
3. If a 2nd network card is not used, close all the applications on the notebook.
 Applications requiring Internet or a network, such as e-mail, SAP applications,
Internet or Windows Explorer.
4. Close any open Internet browsers.
5. Configure the properties of the Internet protocol (TCP/IP) as defined in the table:
IP address
192.168.1.XXX; for XXX all numerical sequences except: 0, 212 and 255 → e.g.
192.168.1.213
Subnet mask
255.255.255.0
Default gateway
192.168.1.212 or leave cells empty
Starting the Web browser
1. Start the Web browser on the computer.
Proline Cubemass C 100 PROFIBUS DP
Endress+Hauser

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Proline Cubemass C 100 PROFIBUS DP and is the answer not in the manual?

Subscribe to Our Youtube Channel

Table of Contents