Power Supply; Terminal Assignment; Available Device Plugs; Supply Voltage - Endress+Hauser Proline Prosonic Flow P 500 Technical Information

Ultrasonic time-of-flight flowmeter
Hide thumbs Also See for Proline Prosonic Flow P 500:
Table of Contents

Advertisement

Proline Prosonic Flow P 500

Terminal assignment

Available device plugs

Endress+Hauser
Supported baud rate
Data transmission mode
Data access
System integration

Power supply

Transmitter: supply voltage, input/outputs
HART

Supply voltage

1 (+)
2 (–)
26 (+)
Modbus RS485
Supply voltage
1 (+)
2 (–)
26 (B)
Transmitter and sensor connection housing: connecting cable
The sensor and transmitter, which are mounted in separate locations, are interconnected by a
connecting cable. The cable is connected via the sensor connection housing and the transmitter
housing.
Terminal assignment and connection of the connecting cable:
Proline 500 →  31
Device plugs may not be used in hazardous areas!
Device plug for connecting to the service interface:
Order code for "Accessory mounted"
Option NB, RJ45 M12 adapter (service interface) →  30
• 1 200 BAUD
• 2 400 BAUD
• 4 800 BAUD
• 9 600 BAUD
• 19 200 BAUD
• 38 400 BAUD
• 57 600 BAUD
• 115 200 BAUD
• ASCII
• RTU
Each device parameter can be accessed via Modbus RS485.
For Modbus register information →  86
Information regarding system integration: Operating Instructions .
• Modbus RS485 information
• Function codes
• Register information
• Response time
• Modbus data map
Input/output
Input/output
1
27 (–)
24 (+)
The terminal assignment depends on the specific device version ordered .
Input/output
Input/output
1
27 (A)
24 (+)
The terminal assignment depends on the specific device version ordered .
Input/output
2
3
25 (–)
22 (+)
Input/output
2
3
25 (–)
22 (+)
23 (–)
23 (–)
29

Advertisement

Table of Contents
loading