Commissioning
Parameter
Assign simulation process variable
Value process variable
Simulation device alarm
Diagnostic event category
Simulation diagnostic event
*
Visibility depends on order options or device settings
76
Navigation
"Diagnostics" menu → Simulation
‣
Simulation
Assign simulation process variable
Value process variable
Simulation device alarm
Simulation diagnostic event
Parameter overview with brief description
Prerequisite
–
A process variable is selected
in the Assign simulation
process variable parameter
(→ 76).
–
–
–
10.7
Protecting settings from unauthorized access
The following options exist for protecting the configuration of the measuring device from
unintentional modification after commissioning:
• Write protection via access code for Web browser → 77
• Write protection via write protection switch → 77
Description
Select a process variable for
the simulation process that is
activated.
Enter the simulation value for
the selected process variable.
Switch the device alarm on and
off.
Select a diagnostic event
category.
Select a diagnostic event for
the simulation process that is
activated.
Proline Cubemass C 100 PROFIBUS DP
Selection / User
Factory setting
entry
• Off
Off
• Mass flow
• Volume flow
• Corrected volume
flow
• Density
• Reference density
• Temperature
*
• Concentration
*
• Target mass flow
*
• Carrier mass flow
Depends on the
0
process variable
selected
• Off
Off
• On
• Sensor
Process
• Electronics
• Configuration
• Process
• Off
Off
• Diagnostic event
picklist (depends
on the category
selected)
Endress+Hauser
→ 76
→ 76
→ 76
→ 76
Need help?
Do you have a question about the Proline Cubemass C 100 PROFIBUS DP and is the answer not in the manual?