Endress+Hauser Proline Cubemass C 100 PROFIBUS DP Operating Instructions Manual page 113

Coriolis flowmeter
Hide thumbs Also See for Proline Cubemass C 100 PROFIBUS DP:
Table of Contents

Advertisement

Proline Cubemass C 100 PROFIBUS DP
Repeatability
Endress+Hauser
Temperature
±0.5 °C ± 0.005 · T °C (±0.9 °F ± 0.003 · (T – 32) °F)
Zero point stability
DN
[mm]
1
2
4
6
Flow values
Flow values as turndown parameter depending on nominal diameter.
SI units
DN
1:1
[mm]
[kg/h]
1
20
2
100
4
450
6
1 000
US units
DN
1:1
[inch]
[lb/min]
¹⁄₂₄
0.735
¹⁄₁₂
3.675
¹⁄₈
16.54
¹⁄₄
36.75
Accuracy of outputs
The output accuracy must be factored into the measured error if analog outputs are
used, but can be ignored for fieldbus outputs (e.g. Modbus RS485, EtherNet/IP).
The outputs have the following base accuracy specifications.
o.r. = of reading; 1 g/cm
3
Base repeatability
Design fundamentals →  115
Mass flow and volume flow (liquids)
±0.05 % o.r.
[in]
¹⁄₂₄
¹⁄₁₂
¹⁄₈
¹⁄₄
1:10
1:20
[kg/h]
[kg/h]
2
1
10
5
45
22.5
100
50
1:10
1:20
[lb/min]
[lb/min]
0.074
0.037
0.368
0.184
1.654
0.827
3.675
1.838
= 1 kg/l; T = medium temperature
Technical data
Zero point stability
[kg/h]
[lb/min]
0.0008
0.00003
0.002
0.00007
0.014
0.0005
0.02
0.0007
1:50
1:100
[kg/h]
[kg/h]
0.4
0.2
2
1
9
4.5
20
10
1:50
1:100
[lb/min]
[lb/min]
0.015
0.007
0.074
0.037
0.331
0.165
0.735
0.368
1:500
[kg/h]
0.04
0.2
0.9
2
1:500
[lb/min]
0.001
0.007
0.033
0.074
113