Endress+Hauser Proline Cubemass C 100 PROFIBUS DP Operating Instructions Manual page 118

Coriolis flowmeter
Hide thumbs Also See for Proline Cubemass C 100 PROFIBUS DP:
Table of Contents

Advertisement

Technical data
Sensor housing
Rupture disk
Flow limit
Pressure loss
System pressure
118
The sensor housing is filled with dry nitrogen gas and protects the electronics and
mechanics inside.
To guarantee the safety of the measuring device, the device version with a rupture disk
with a triggering pressure of 10 to 15 bar (145 to 217.5 psi) is the standard version used.
Special mounting instructions →  22.
Select the nominal diameter by optimizing between the required flow range and
permissible pressure loss.
For an overview of the full scale values for the measuring range, see the "Measuring
range" section →  108
• The minimum recommended full scale value is approx. 1/20 of the maximum full scale
value
• In most applications, 20 to 50 % of the maximum full scale value can be considered ideal
• A low full scale value must be selected for abrasive media (such as liquids with entrained
solids): flow velocity < 1 m/s (< 3 ft/s).
• For gas measurement the following rules apply:
• The flow velocity in the measuring tubes should not exceed half the sound velocity
(0.5 Mach).
• The maximum mass flow depends on the density of the gas: formula →  108
To calculate the flow limit, use the Applicator sizing tool →  106
To calculate the pressure loss, use the Applicator sizing tool →  106