11.2 Evaluate selftest file

Path: C:\ProgramData\Rohde-Schwarz\TSMA\Selftestresults
Filename: Selftestresult.txt
Entries in the Selftestresult.txt are created during the boot process of the
device and every 5 minutes.
To open the Selftestresult.txt, proceed as follows.
1. Open a file explorer.
2. Enter the path to the directory where the selftest file is stored.
3. Open the selftest file.
Following device parameters and values are stored.
Internal voltages (nominal values / actual values)
Temperatures (scanner / mainboard)
Battery status info
Status of fans
Status of scanner interface

11.3 Access the firmware log file

The firmware log file is created once per day. A maximum of 15 files are available for
download (see
To check the log file entries
1. Navigate to "File Transfer" > "App Log".
2. Under "Firmware Log Trace", you see the content of the current log file.
To download a log file
1. Navigate to "File Transfer" > "App Log".
2. Under "Firmware Log File", select the log file according to your needs.
The log files have the following format: Log_yyyy-mm-dd.txt.

11.4 Write battery info into firmware log file

As an option, you can write the battery info additionally into the firmware log file (see
"Battery"
User Manual 4900.8057.02 ─ 11
"App Log"
on page 166).
on page 159.
Troubleshooting

Write battery info into firmware log file