8.2 Evaluate Selftest File

Path: C:\ProgramData\Rohde-Schwarz\TSMA\Selftestresults
Filename: Selftestresult.txt
Entries in the Selftestresult.txt are created during the boot process of the
device and every 5 minutes.
To open the Selftestresult.txt, proceed as follows.
1. Open a file explorer.
2. Enter the path to the directory where the selftest file is stored.
3. Open the selftest file.
Following device parameters and values are stored.
Internal voltages (nominal values / actual values)
Temperatures (scanner / mainboard)
Battery status info
Status of fans
Status of scanner interface
8.3 Verify Scanner Link / Recall Device Info Using Tsme-
DeviceManager
1. To start the R&S TSME Device Manager, double-click the desktop icon or select in
Windows:
"Start" > "All Programs" > "TsmeTools" > "TsmeDeviceManager"
2. Following error messages may occur.
3. If the"Device Info" tab (see
tion has been established successfully.
4. Check the "Device Analysis Output" (see
on page 96) and the "Options" (see
Keys",
User Manual 4900.8057.02 ─ 06

Verify Scanner Link / Recall Device Info Using TsmeDeviceManager

Tsme Device Manager Application has stopped
In the web GUI. check "Mode of Operation". If necessary, change to "PC Mode"
and try again.
No TSME connected
In the web GUI, verify "Scanner LAN Port Settings".
on page 97) .
Figure
8-1) contains scanner data, the scanner connec-
Chapter 8.3.2, "Device Analysis
Chapter 8.3.3, "Verify Installed License
Troubleshooting
Output",