Accumulator - LINET Multicare X Instructions For Use And Technical Description

Positionable bed for intensive care version with scales
Table of Contents

Advertisement

10 Accumulator

WARNING!
When the bed is not connected to the mains and accumulator is not sufficiently charged all electrical func-
tions of the bed are blocked! After some long period without charging the battery, the battery will be deeply
discharged and loose its full functionality!
Purpose
The accumulator serves as a backup during power failures or for emergency bed positioning.
Use only accumulator approved by the manufacturer.
Check the functionality of the accumulator at least once a month and have the accumulator changed if necessary.
The manufacturer will assume no responsibility for any damage to the bed or the accumulator caused by:
non-observance of the manufacturer's instructions in the instructions for use.
using accumulator not approved by the manufacturer.
Warranty
The manufacturer provides a 6-month warranty for the full function of the accumulator.
Accumulator lifetime could be up to 5 years if operated under optimum conditions.
Accumulator capacity can be significantly reduced if:
too high ambient temperature
many accumulator charge/discharge cycles
recurrence of deep discharge
bed is often powered only by the accumulator
Charging
The accumulator supplied with the bed is delivered insufficiently charged.
Charging of the accumulator before use of the bed takes approximately 4 hours.
Charging of the accumulator runs automatically when the power cable is connected to the mains.
To charge the accumulator:
Connect the bed to the mains.
Storage
For declared lifetime period of leaded accumulators is recommended during storage:
To prevent accumulator from deep discharging and to keep accumulator at least partly charged by regular recharging.
To store accumulator on the dry places with temperature from 10°C to 40°C.
To prevent accumulators from being in the sunshine.
Signalisation
The indicator on Multiboard X touchscreen and LED on Attendant Control Panel indicate the accumulator charge status.
D9U001MCX-0101_03