  • Page 1 User manual and technical description Eleganza Smart Junior D9U001GMR-0101 Version: 01 Publication Date: 2017-06...
  • Page 2 Translation © LINET, 2017 All rights reserved. All trademarks or names are the property of the respective owners. LINET, s.r.o. reserves the right to change the specifications at any time without warning. The information included in this document is presented by LINET in an attempt to accurately and correctly inform the customer. However LINET shall not be liable for the consequences arising from the use of this information, the infringement of patent or other rights of third parties that arises from the use of this information.
  • Page 3 Dear Customers, We thank you for deciding to purchase our company’s product and hope that you find it satisfactory throughout its service life. Thanks to the very careful selection of materials, the latest production technology and careful work of our employees, our bed meets the parameters that guarantee its high quality, reliability and constant utility value.
  • Page 4: Table Of Contents

    Table of contents: Introduction ........................6 Method of Delivery ........................7 Safety Instructions ......................7 Conditions of Use ......................9 Technical Parameters ....................10 Variant Specification ....................11 Bed Description ......................12 Bed Installation and Putting Into Operation ............... 13 Connecting the Bed to the Mains ....................13 Connecting to Adjust the Potentials ....................13 Charging the Backup Battery .....................14...
  • Page 5 Setting the Battery Quiet Mode ....................31 Maintenance........................ 31 Environmental protection ................... 32 Guarantee and Service ....................33 Signs and Symbols Used .................... 34 List of Service Procedures and Regular Service Maintenance........35 Eleganza Smart Junior Bed Completion Certificate ..........36 Serial Label ......................... 36 D9U001GMR-0101_01...
  • Page 6: Introduction

    Introduction The adjustable Eleganza Smart Junior bed is designed for the hospitalisation of paediatric patients in the age of 4 - 15 years with minimal height of 90cm, on standard hospital and long-term care wards. The manufacturer does not exclude placing the bed in intensive care units, but this can be considered only on the basis of the responsible choice of the medical care provider.
  • Page 7: Method Of Delivery

    The castors may be damaged when travelling with the bed on a rough, uneven and dirty surface. Safety Instructions Eleganza Smart Junior bed should be left in its lowest position when the patient is unattended in order to reduce risk of injury due to falls! WARNING! Siderails of Eleganza Smart Junior should be located in the „up“...
  • Page 8  Only Linet mattresses and accessories may be used on the bed; if incompatible mattresses are used this may result in the bad function of the siderails and endanger the safety of patients ...
  • Page 9: Conditions Of Use

     Undercarriage of the bed is not intended for use as a storage place. In case of placing any objects to undercarriage and to the space under the mattress platform, during the height adjustment there can happen a damage of bed or these objects. ...
  • Page 10: Technical Parameters

    Technical Parameters Bed Technical Parameters Description Value Outer dimensions 189,7 x 92,1 cm Bed with 1-part collapsible siderails Recommended mattress dimension – standard 170 × 80 cm mattress platform dimension Max. mattress height 14 cm 38,5 cm – 76 cm Mattress platform height Backrest –...
  • Page 11: Variant Specification

    Variant Specification Variable feature (equipment) - possibility of choice for the customer Basic bed variant  Standard bed dimensions Bed mattress platform  Mattress platform made of metal segments Siderails  Bed without siderails  Single-section collapsible siderails with control on the bottom Ends ...
  • Page 12: Bed Description

    Bed Description 1. Bed foot end 2. Mattress platform seat rest (solid) 3. Single-section coated collapsible siderails 4. Bed head end 5. Mattress platform backrest 6. Lifting pole, infusion stand or other accessory holder 7. CPR control lever (cardio-pulmonary resuscitation) unblocking the backrest 8.
  • Page 13: Bed Installation And Putting Into Operation

    Bed Installation and Putting Into Operation Remove the packaging from the bed and destroy the ecological packaging material. If the beds are delivered in double packs (two beds one above the other), carefully remove the top bed and lay it on the floor. At least two persons must carry this out! Check the delivery is complete according to the delivery note.
  • Page 14: Charging The Backup Battery

    7.3 Charging the Backup Battery The bed may be delivered with an attached backup battery. Before starting to use the bed the battery must be charged to the sufficient capacity by connecting the bed to the mains. The course of the charging and capacity of the battery is indicated on the nurse control panel with the aid of a yellow warning light.
  • Page 15: Bed Adjustment

    Bed Adjustment The various controls stated in the next chapters can be used to electrically adjust the bed. The choice of controls corresponds to the bed equipment ordered by the customer. 8.1 Nurse Control Panel (Supervisory Panel) The nursing staff control panel is kept in the rack under the foot end. The control can be removed and can be placed on the bed end or hand held when controlling the bed.
  • Page 16: Activating "Go" Button

    1. Central “STOP“ button 2. Activating “GO“ button 3. Central locking button for individual functions 4. Warning light for the thigh rest lock 5. Button for adjusting the mattress platform thigh rest 6. Warning light for the backrest lock 7. Buttons for adjusting the mattress platform backrest 8.
  • Page 17 8.1.3.1 Resuscitation Position – CPR This safety position is designed for resuscitation. The individual parts of the mattress platform and the mattress platform itself are set at the horizontally lowest position. The keys must be held until the position is finally set. This function cannot be blocked. 8.1.3.2 Trendelenburg Anti-shock Position This safety position is used for a situation when anti-shock conditions must be secured for a person on the bed.
  • Page 18 8.1.4 Locking the Individual Positions The bed is equipped with a safety function disabling the setting of the individual positions by using locks on the central control panel. The blocking is carried out in the same way as with standard positioning: 1.
  • Page 19: Hand Control

    Hand Control The hand control is used by the patient, the operator respectively, to adjust the individual parts of the mattress platform of the bed. The same procedure applies for setting the positions on the hand control as when adjusting the bed’s positions with the aid of the nurse control panel: activating the keyboard by pressing the activating “GO”...
  • Page 20: Cpr Unblocking Of Backrest

    CPR Unblocking of Backrest urgent situations such as immediate reanimation (CPR) the bed allows the backrest to be quickly mechanically folded. The CPR lever is used for this purpose located under the bed frame at the head of the patient. This method of folding the backrest is used particularly when there is a power cut.
  • Page 21: Calf Rest

    Calf Rest The position of the calf rest can be set mechanically with the aid of a serrated raster using the procedure described hereunder. By gripping both handrails of the rest and continuously pulling, the calf rest can be set to the appropriate position. If the calf rest needs to be lowered from a higher to a lower position, the arrestment mechanism has to be released firstly by slightly pulling it upwards and then continuously lowering the calf rest...
  • Page 22: Bed Ends

     Bed Ends The Eleganza Smart Junior beds are equipped with two removable (head and foot) ends, which are offered in several differend materials used. The used material does not affect the end removal procedure. Both ends can be locked so they cannot be pulled out by a safety lever in the relevant position.
  • Page 23: Siderails

    The Eleganza Smart Junior beds are delivered wither without siderails or with single-section collapsible siderails (1 pair per bed). When you order Eleganza Smart Junior bed without any side-rails, it is not possible to order the side-rails later for this bed. The construction of the bed does not allow subsequent assembly of the side-rails.
  • Page 24: Central Braking System

    By pulling the siderail in the opposite direction to an erect position in which the arrestment mechanism will be secured spontaneously. Do not use part of the siderails marked on figure (picture below) for handling the siderail – in certain situations the operator may trap his fingers between the siderails and the adjustable backrest or head end of the bed.
  • Page 25: Accessories

    The appropriate height can be adjusted for the patient by shortening this belt. Linet is offering plastic triangular handrails, telescopic infusion stands and stainless steel baskets for infusion solutions suitable for Eleganza Smart Junior bed. The infusion stand can be slid into the opening at the head end together with 4 hooks for suspending the bottle or basket for the infusion solutions.
  • Page 26: Small Accessory Positioning Holder

    The lfiting pole weighs 75 kg and is not designed for the rehabilitation exercises of patients. 16.2 Small Accessory Positioning Holder The Eleganza Smart Junior bed may be equipped for an extra charge with two holders for various types of accessories.
  • Page 27: Other Accessories

    16.3 Other Accessories Towel holder coated This is suitable for use only on plastic ends – it cannot be used on ends with formica panelling. The method of location and use is clear from the figure. It is made from ABS plastic, is suspended Writing rack through openings in ABS ends and when travelling with the bed or for a different...
  • Page 28: Cleaning And Disinfection

    Cleaning and Disinfection 17.1 Basic Instructions Before You Start Cleaning Use disinfectants designed only for cleaning healthcare equipment. Do not use abrasives (scouring powder), scourers and other materials and agents which could damage the coating. Do not use corrosive and caustic agents and strong acids. Do not use agents which could alter the structure and behaviour of the plastics (petrol, toluene, acetone, etc.).
  • Page 29: Extent Of Cleaning

    17.3 Extent of Cleaning Table No.4 — Extent of Cleaning Extent of decontamination – bed parts Level of cleaning 1. Daily cleaning Surfaces exposed to daily contact with the patient and operator: - All control features for bed adjustment - The entire area of the ends including the side columns and parts under both ends - The entire area of the siderails (in the exposed top position) - Freely accessible mattress surface...
  • Page 30 4. Control of one of the controls does The bed is not Check the connection to the not work; the height and the connected to the mains. position of the parts of mattress mains. platform cannot be adjusted. Check and unblock the locks on 5.
  • Page 31: Storage

    Storage For problem-free storage it is recommended: - Disconnect the bed from the electric mains feed and set the “Quiet Mode” of the backup battery on the central control panel (see next chapter). - Dismantle the accessories (straightening lifting pole etc.) - Wrap or cover the bed and accessories in such a way so that the coating and plastic parts cannot get damaged.
  • Page 32: Environmental Protection

    21 Environmental protection The company LINET, spol. s r.o. is aware of the necessity to protect the environment for future generations and therefore devotes great attention to the development, innovation – planning, production and use of such technology and materials that are environmentally friendly.
  • Page 33: Guarantee And Service

    By removing this product appropriately you will help preserve valuable natural resources and help prevent potential negative effects on the environment and human health, which could result from the incorrect destruction of waste. You can request further details from authorised environmental protection authorities, or the closest collection point for separated waste collection.
  • Page 34: Signs And Symbols Used

    23 Signs and Symbols Used Transformer thermal protection Warning in the instruction manual Instrument for indoor use Applied Parts – type B Electromagnetic compatibility D9U001GMR-0101_01...
  • Page 35: List Of Service Procedures And Regular Service Maintenance

    24 List of Service Procedures and Regular Service Maintenance PROCEDURE PURPOSE AND DESCRIPTION DATE PROCEDURE CARRIED OUT BY D9U001GMR-0101_01...
  • Page 36: Eleganza Smart Junior Bed Completion Certificate

    25 Eleganza Smart Junior Bed Completion Certificate Order number: customer: model number: supplier: date of acceptance: serial number: I confirm that the operation staff has been trained to learn to use the bed. Guarantee period: ........................date signature and stamp of the customer signature and stamp of the supplier This completion certificate also serves as a guarantee certificate.
  • Page 37 D9U001GMR-0101_01...