19.4 Infusion Stand
WARNING!
Risk of injury due to use of incorrect accessories or because of incorrect use!
Infusion Stands must only be used for their intended use. Always read the instructions for use of the corre-
sponding infusion stand!
►
Only mount an infusion pump to the lower (wider) telescopic section of an infusion stand above the head board or
foot board.
►
Never mount an infusion pump to the upper (thiner) telescopic section of an infusion stand.
►
Ensure the infusion pump will not collide with any movable parts of the bed (especially Backrest part) or with the
patient. This must be verified during installation.
►
Do not over tighten the infusion pump clamps during fitment. Overtightening may damage the infusion stand.
►
Infusion pump can be only used if the infusion stand is fitted in the accessory holder socket in the head end on
the under carriage of the bed.
►
Do not use the infusion stand as driving/pushing device during the bed transport.
►
Do not insert Infusion Stand 4PKV26107200 to the holes in the corner bumpers!
►
Use exclusively infusion stands with hooks for hanging IV bags or baskets for intravenous solutions.
►
Ensure the infusion stand individual hook 2kg maximum Safe Working Load is not exceeded.
►
Ensure the infusion stand 20kg maximum Safe Working Load is not exceeded.
Infusion stands are intended to provide a suitable support for the attachment of the infusion pumps/syringe pumps and suspension
of the IV bags, bottles or baskets for intravenous solutions.
Fig. Places for Infusion Stand 4PKV26107200 (sleeve fittings on
accessory adaptor)
D9U001MCX-0101_03
Fig. Infusion Stand with
receptacle for cannulas
(4PKV26107200)
145