Status (Pop-up window)
MAX 15°
Transfer mode active.
Lateral tilt limited to 15°.
Max angle by foot controls reached.
Use controls in siderail or ACP to
continue.
MAX 15°
Lateral tilt limited to 15°.
Lower the angle of backrest bellow 30°
to continue.
MAX
Lateral tilt is limited.
Load exceeded.
TR/ATR locked during lateral tilt.
ALT disabled. Bed disconnected.
Plug bed into socket.
52
Lateral Tilt with siderail folded down is
limited to 15°. Transfer Mode is activated.
Lateral Tilt that can be adjusted by pedals
of Lateral Tilt Foot Control is limited to
Lateral Tilt is limited to 15° because Bac-
>30°
Lateral Tilt is limited to 15° because the
bed is loaded with more than 150 kg.
Longitudinal tilt adjustment (Trendelen-
burg tilt and Antitrendelenburg tilt) is disa-
bled during the set lateral tilt of the bed.
Automatic Lateral Therapy is disabled
because the bed is disconnected from
Meaning
15°.
krest Angle is in more than 30°.
the mains power.
How to change the status
Use Lateral Tilt with siderail
folded down to facilitate patient
transfer.
Use LCD touchscreen on Mul-
tiboard X or Attendant Control
Panel to continue in lateral tilting
above 15°.
Lower the Beckrest to less than
30° to continue in lateral tilting
above 15°.
Remove load to continue in late-
ral tilting above 15°.
Decrease the lateral tilt angle to
continue in longitudinal tilt adjus-
tment.
Connect bed to the mains power
to enable the Automatic Lateral
Therapy.
D9U001MCX-0101_03