NEW PATIENT GUIDE
Order of the NEW PATIENT Guide steps
1)
Ensure the following conditions are met:
►
Patient is not in bed.
►
Scales are stabilized.
To start the preparation of the bed for new patient:
►
Press Tick Icon.
To cancel the NEW PATIENT function:
►
Press Cross Icon.
PREPARE THE BED FOR NEW PATIENT
2)
To continue preparing the bed for new patient:
• Make the bed
►
• Attach the required accessories
Make the bed.
• Set the bed to horizontal position
►
Add the required accessories to the bed.
►
Adjust the bed to horizontal position by pressing the
icon
.
It is not possible to advance to the next step without the bed in
horizontal position!
►
Press Tick Icon.
3)
Do not touch the bed and wait for the end of the proce-
ssing!
4)
To fi nish the preparation of the bed for new patient:
Press Tick Icon.
►
The bed is prepared for new patient's admission.
64
Signalisation & Settings
NEW PATIENT – RESET BED DATA
• The patient can not be in bed
• Tare/Zero scales
• Reset all patient histories
• Reset mattress setting
PREPARE THE BED FOR NEW PATIENT
• Make the bed
• Attach the required accessories
• Set the bed to horizontal position
PROCESS IN PROGRESS
Do not touch the bed!
BED IS READY FOR PATIENT
Reset of scales, histories and
mattress setting
D9U001MCX-0101_03