New Patient Function - LINET Multicare X Instructions For Use And Technical Description

Positionable bed for intensive care version with scales
Table of Contents

Advertisement

SERVICE SETTINGS
Fig. Settings Screen 3 (insert password to enter Service Settings)

12.1.5 NEW PATIENT FUNCTION

It is recommended to use New Patient function when replacing patients and no patient lies on the Multicare X bed.
The New Patient function has following eff   ects on the Multicare X bed with OptiCare X integrated mattress:
Scales Taring (only if the scales are stabilised)
Deletion of the Scales history
Deletion of the ALT history
Deletion of the Backrest History
Setting of the MOBILE Mode with pressure level 3 (only for OptiCare X)
Deletion of the Pressure Comfort Setting (only for OptiCare X)
Setting of the MCM with HIGH intensity level (only for OptiCare X)
To activate New Patient function:
Wait until the scales are stabilized.
Press and hold NEW PATIENT Icon.
Follow NEW PATIENT Guide.
Beep indicates confi rmation of the activated New Patient function.
Scales have been tared, histories deleted and it is recommended to place new patient on the bed.
D9U001MCX-0101_03
PASSWORD
* * * * 6
1
2
3
4
5
6
7
8
9
0
NEW PATIENT