15.2 Activation

Fig. Bed Exit Alarm Monitoring Screen (OFF)
To enter Bed Exit Alarm Monitoring Screen:
Press Bed Exit Alarm Monitoring Screen Icon (1).
To activate Bed Exit Alarm Monitoring:
Press ON Icon (2) when patient is on the bed.
Fig. Notifi cation of Insuffi ciently Loaded Bed
D9U001MCX-0101_03
01
BED EXIT IS
OFF
2
06
OFF
< MIN
BEA cannot be activated
- Load is <35 kg.
BED EXIT IS
OFF
OFF
1
1.
Bed Exit Alarm Monitoring Screen
Icon
2.
ON Icon (activation)
22:58
25. 12. 25
22:58
25. 12. 25
Bed Exit Alarm Mo-
nitoring Screen is
displayed with Bed
Picture and Patient
Picture.
If you press ON
Icon without patient
on the bed, Notifi -
cation of Insuffi ci-
ently Loaded Bed is
displayed and the
Bed Exit Monitoring
is not activated.
Minimum patient
weight for Bed Exit
Monitoring is 35 kg.