21 Troubleshooting
DANGER!
Risk of mortal injury due to electric shock!
►
If a fault occurs, have the electric motor, power box or other electrical parts repaired by qualified personnel exclu-
sively.
►
Do not open the protective covers of the electric motor or the power box.
Error/Fault
Adjusting with position buttons not
possible
Faulty Mattress support platform
Height/Tilt Adjustment
Lowering Backrest from the upright
position not possible
Adjusting Siderails not possible
Faulty brakes
166
Cause
GO Button was not pressed.
Function disabled on Attendant Control Panel.
Actuators have no power.
Defective actuators.
Defective accumulator.
Mains Plug inserted incorrectly.
Faulty Power Source.
Faulty Control Element.
Obstacle on the undercarriage cover.
Function disabled on Attendant Control Panel.
Actuators have no power.
Defective actuators.
Defective accumulator.
Mains Plug inserted incorrectly.
Faulty Power Source.
Faulty Control Element.
Obstacle under the Backrest or in the drive
mechanism.
CPR Release Handle is defective.
Obstacle in the Siderail Release Mechanism.
Siderail Release Mechanism is defective.
Obstacle blocking brakes mechanically.
The brake mechanism is defective.
Solution
Press the GO button.
Enable disabled function.
Check the mains connection.
Notify the manufacturer´s service
department.
Insert the Mains Plug correctly.
Notify the manufacturer´s service
department.
Notify the manufacturer´s service
department.
Remove the obstacle.
Enable disabled function.
Check the mains connection.
Notify the manufacturer´s service
department.
Insert the Mains Plug correctly.
Notify the manufacturer´s service
department.
Notify the manufacturer´s service
department.
Remove the obstacle
Notify the manufacturer´s service
department.
Remove the obstacle.
Notify the manufacturer´s service
department.
Remove the obstacle.
Notify the manufacturer´s service
department.