15.6 BED EXIT ALARM

Audible alarm is triggered when patient has left selected monitored zone or when PAUSE period is terminated and patient is not
just in the ordered position. During this alarm a text „BED EXIT ALARM" in a red rectangle is displayed on the Bed Exit Alarm Moni-
toring Screen.
To stop Alarm:
Press OFF Icon (5).
Bed Exit Monitoring is deactivated and text „BED EXIT IS OFF" in a yellow fi eld is displayed on the screen.
The audible alarm is muted.
To pause Alarm:
Press PAUSE Icon (6).
Countdown timer (15 min) appears on the display in the upper right corner of the screen. The audible alarm is muted.
ZONE
Fig. BED EXIT ALARM (Alert)
ALARM VOLUME
It is possible to set Alarm Volume before and during triggered audible alarm.
Maximum Alarm Volume Level is set by default.
It is not possible to mute completely the audible alarm with this volume setting.
To lower Alarm Volume Level:
Press Volume Icon (7).
Indicator with lower volume levels (8) appears on the display.
Volume is lowered.
To return to Maximum Alarm Volume Level:
Press Volume Icon (7) after Minimum Alarm Volume Level has been reached.
Indicator with the 3 volume levels (8) appears on the display.