For safe and gentle cleaning:
►
Do not use any strong acids or bases (optimum pH range 6 - 8).
►
Exclusively use detergents that are suitable for cleaning medical equipment.
►
Do not use abrasive powders, steel wool, or other materials and cleaning agents that might damage the mattress replace-
ment system.
►
Never use any corrosive or caustic detergents.
►
Never use detergents that deposit calcium carbonate.
►
Never use detergents with solvents that might affect the structure and consistency of the plastics (benzene, toluene, ace-
tone, etc.).
►
Clean electrical components carefully and allow them to dry completely.
►
Do not immerse SCU in water or steam-clean it.
►
Observe local directives regarding infection control.
►
Make sure any cleaning agent used is approved by:
■
the facility in which the mattress replacement system is to be used.
■
by the environmental protection agency of the country in which the mattress replacement system is to be used.
20.1 Cleaning (Multicare X)
Prepare for cleaning as follows:
►
Put the mattress support platform in the highest position.
►
Adjust the back and thigh rests so that the reverse sides are accessible.
►
Disable the function buttons on the control elements using the Attendant Control Panel.
►
Disable the foot controls using the Attendant Control Panel.
►
Disconnect the bed from the mains.
►
Move the bed to the location where it will be cleaned.
►
Lock the brakes on the bed.
20.1.1 Daily Cleaning
Clean the following bed parts:
■
All control elements for adjusting the bed
■
All handles
□
CPR release handle
■
Head board and Foot board
■
Siderails (in highest position)
■
Freely accessible mattress surface
■
Mobi-Lift
®
■
Accessory rails
20.1.2 Cleaning before Changing Patients
Clean the following bed parts:
■
All control elements for adjusting the bed
■
All handles
□
CPR release handle
■
Head board and Foot board
■
Siderails (in highest position)
■
Freely accessible mattress surface
■
Mobi-Lift
®
■
Accessory rails
■
All plastic mattress support platform covers
■
Plastic undercarriage covers
■
Telescopic columns
■
Mattress on all sides
■
Freely accessible metal parts of mattress support platform
■
Cable ducts
■
Lifting pole sleeve fitting
■
Infusion stand sleeve fitting
■
Bumpers
■
Castors
■
Brakes