Before Use; Transport; Bed Transport Mode - LINET Multicare X Instructions For Use And Technical Description

Positionable bed for intensive care version with scales
Table of Contents

Advertisement

8.6 Before Use

Prepare the bed for use as follows:
Connect the bed to the mains.
Charge the accumulator.
Raise and tilt the mattress support platform to the highest position.
Lower and tilt the mattress support platform to the lowest position.
Check that the castors as well as main brake work correctly.
Check that the bed extension works correctly.
Check that it is possible to remove the head board and foot board.
Check all of the functions on the control elements.
Check that the siderails function properly.
Dispose of all packaging (see Disposal).

8.7 Transport

For a safe transport, observe the following:
Ensure that no cables are run over when moving a bed.
Ensure that the mains cable is attached with a hook (at the head end of the bed).
Ensure that the castors are unlocked before moving the bed during the loading/unloading process (see Castor Control).
Adjust bed height to at least 20 cm below maximum height.
Push bed by handles on head board or foot board.
Move the bed exclusively on suitable floor surfaces.
Ensure the bed is braked when it is not transported.
For longer distances, ensure that the castor steering function (main control) is activated.
Ensure that the brakes are released while moving the bed.
Suitable surfaces:
Tile
Hard linoleum
Poured flooring
Unsuitable surfaces:
Too soft, unsealed or defective flooring
Soft wooden flooring
Soft and porous stone floors
Carpeted floors with underlay
Soft linoleum

8.7.1 Bed Transport Mode

To avoid damaging the bed and the environment during transport or storage:
Activate Bed Transport Mode on the Attendant Control Panel.
To activate Bed Transport Mode:
Disconnect the bed from the mains.
Disconnect the ground wire.
Activate the keypad by pressing the button
Press the Thighrest Up button + Thighrest Down button + Trendelenburg Tilt button at the same time and hold them for
three seconds.
The Bed Transport Mode is activated.
To deactivate Bed Transport Mode:
Connect Power Cable to the mains.