10.2 Potential Interconnection (optional)
The stretcher is equipped with a standard protective connector. This connector is used for potential equalisation between the stret-
cher and any intravascular or intracardiac device connected to the patient to protect the patient from static electric shocks.
Fig. Potential equalisation - male (head end, bottom view)
Use equalisation connector if:
■
the patient is connected to any intravascular or intracardiac device.
Before connecting the patient to an intravascular/intracardiac device:
►
Connect the ground wire of the device to the potential equalisation
connector on the stretcher on which the patient in question is lying.
►
Use a standard hospital connector.
►
Make sure that the connectors match.
►
Make sure that there is no possibility for inadvertent disconnection.
Before moving the stretcher:
►
Disconnect the patient from the intravascular or intracardiac device.
►
Disconnect the potential equalisation connector.
10.3 Before Use
Prepare the stretcher for use as follows:
►
Perform bleeding procedure of hydraulic units in the highest stretcher position by pressing lifting pedal 10 times.
►
Raise the Mattress Support Platform to the highest position.
►
Lower the Mattress Support Platform to the lowest position.
►
Tilt the Mattress Support Platform to Trendelenburg position.
►
Position the Mattress Support Platform to Anti-Trendelenburg tilt.
►
Check that the castors as well as main brake work correctly.
►
Check that the Siderails function properly.
►
Dispose of all packaging (see „25 Disposal (Sprint 200 without scales and without i-Drive Power)" on the page 134 or see
„26 Disposal (Sprint 200 with scales or with i-Drive Power)" on the page 135).
D9U001ES2-0101_05
Fig. Potential equalisation connector -
female
43