Output Filtering Option; Select Function Option - Ametek Thermox CG1000 User Manual

Table of Contents

Advertisement

The Series 2000 control unit prompts you with the Set Current Range menu option
where you can choose the range of readings that the 0 to 20 mA or 4 to 20 mA analog
output represents (see the Set Current Range section in this chapter for details).
4-18 | Thermox CG1000 Oxygen Analyzer

Output Filtering Option

Filter out quick transitions in readings to allow a smoother analog output.
Setting output filtering
1. After selecting an analog output port, from the Analog Range Key
select Output Filtering.
2. Enter an output filtering value between 1 and 100 where XX is the cur-
rent filter value, then press Enter.
Note that 100 equals no filtering, and 1 equals the highest possible
filter value. Note that you should enter 01 to enter 1; 07 to enter 7; 70
for 70, etc.
The default value is 100 (no filtering).
NOTE

Select Function Option

Select what type of information the analog output port should track. Choic-
es are:
Oxygen
Cell temperature
Thermocouple millivolts (T/C mV)
Cell millivolts (cell mV)
Selecting the function
1. After selecting an analog output port, from the Analog Range menu,
choose Select Function.
2. Select what you want the analog output port to track.
[100 = FAST
1 = SLOW]
Fltr = XX New #?

Advertisement

Table of Contents
loading

Table of Contents