[:SENSe]:CORRection:TYPE STANdard|FLEX
[:SENSe]:CORRection:TYPE?
Cmd Description: Command sets correction type.
Query Description: Query returns correction type. Query returns STAN|FLEX.
Param Range: STANdard|FLEX
Default Value: STANdard
Example: To set correction type to flex:
[:SENSe]:CORRection[:STATe] ON|OFF|1|0
[:SENSe]:CORRection[:STATe]?
Cmd Description: Command sets correction status.
Query Description: Query returns correction status. Query returns 1 if RF cal is on.
Param Range: ON|OFF|1|0
Default Value: OFF|0
Example: To turn calibration correction on:
[:SENSe]:FREQuency:CABLe <cable-list-index>
[:SENSe]:FREQuency:CABLe?
Cmd Description: Command selects a cable with the index number from the sorted cable
Query Description: Query returns the index number of the selected cable based on the
Param Type: Integer
Param Description: Index number from cable list
Default Value: 0
Example: To set the cable index to 10:
[:SENSe]:FREQuency:CABLe:CATalog:FULL?
Query Description: Query returns the entire cable list that is stored in memory.
Example: To get the cable list:
4-26
Chapter 4 — Cable & Antenna Analyzer Mode Commands
:CORR:TYPE FLEX
Otherwise, it returns 0.
:CORR:STAT ON
list.
sorted cable list.
:FREQ:CABL 10
:FREQ:CABL:CAT:FULL?