:CONFigure:MEASure:DISPlay SINGle|DUAL
:CONFigure:MEASure:DISPlay?
Cmd Description: Command sets the display format to single or dual traces. Only
Query Description: Query returns the display format. Only supported in Advanced Cable &
Param Range: SINGle|DUAL
Default Value: SINGle
Example: To set display format to dual traces:
:CONFigure:MEASure:MODe
RLFReq|RLDTf|SWRFreq|SWRDtf|CLFReq|TR2P|TRES|SMITh|PHASe
:CONFigure:MEASure:MODe?
Cmd Description: Command sets the active trace measurement mode.
Query Description: Query returns the current measurement mode.
Param Description: RLFReq: Return Loss(Frequency domain sweep); RLDTf: Return
Example: To set measurement mode to Cable Loss:
:DISPlay:UPDate ON|OFF|1|0
:DISPlay:UPDate?
Cmd Description: Command sets the display update on/off.
Query Description: Query returns display update status. Query returns 1 if display update
Param Range: ON|OFF|1|0
Default Value: OFF|0
Example: To turn display update off:
4-18
Chapter 4 — Cable & Antenna Analyzer Mode Commands
supported in Advanced Cable & Antenna Analyzer mode.
Antenna Analyzer mode.
:CONF:MEAS:DISP DUAL
Loss(DTF); SWRFreq: SWR(Frequency domain sweep); SWRDtf:
SWR(DTF); CLFReq: Cable Loss - One Port(Frequency domain sweep);
TR2P: Transmission - Two Port (not available in S331L); TRES:
Transmission - External Sensor; SMITh: Smith Chart; PHASe:
One-Port Phase.
:CONF:MEAS:MOD CLFR
is on. Otherwise, it returns 0.
:DISP:UPD OFF
PN: 10580-00322 Rev. F
S331L PM
Need help?
Do you have a question about the Site Master S331L and is the answer not in the manual?
Questions and answers