:CALCulate#:LIMit:UPPer:SEGMent:EDIT
<StartX><StopX><StartY><StopY>
Cmd Description: Command edits the upper limit active segment. The active segment
Param Type: Float
Default Units: X units: Hz for Frequency domain, m or ft for Distance domain; Y units:
Example: To edit the active segment and set it to 1MHz to 1GHz with value 20 dB
Notes: Replace '#' with the required trace number. Check User Guide for the
:CALCulate#:LIMit:UPPer:SEGMent:STARt:X <value>

:CALCulate#:LIMit:UPPer:SEGMent:STARt:X?

Cmd Description: Command sets the upper limit line X start value for active segment.
Query Description: Query returns the upper limit line X start value for active segment.
Param Type: Float
Default Units: Hz for Frequency domain, m or ft for Distance domain, s for Time
Example: To set limit X start value to 1 MHz for trace 1:
Notes: Replace '#' with the required trace number. Check User Guide for the
:CALCulate#:LIMit:UPPer:SEGMent:STARt:Y <amplitude>

:CALCulate#:LIMit:UPPer:SEGMent:STARt:Y?

Cmd Description: Command sets the upper limit line Y start value for active segment.
Query Description: Query returns the upper limit line Y start value for active segment.
Param Type: Float
Default Units: Current active trace y-axis units
Example: To set limit start amplitude to 20 dB for trace 1:
Notes: Replace '#' with the required trace number. Check User Guide for the
4-6
Chapter 4 — Cable & Antenna Analyzer Mode Commands
index must be updated when adding/deleting a segment.
same as current active trace y-axis units
for trace 1:
:CALC1:LIM:UPP:SEGM:EDIT 1e6,1e9,20,20
number of supported traces in each mode.
Time units are available only if option 2 is installed.
domain
:CALC1:LIM:UPP:SEGM:STAR:X 1000000
number of supported traces in each mode.
:CALC1:LIM:UPP:SEGM:STAR:Y 20
number of supported traces in each mode.