Chapter 4 — Cable & Antenna Analyzer Mode Commands
[:SENSe]:CORRection:COLLect:STATus?
{INITialize|OPEN|SHORt|LOAD|SAVE|ALL|THRU|ZERO}
Query Description: Query with step parameter returns 1 if a specified cal step is completed.
Param Range: INITialize|OPEN|SHORt|LOAD|SAVE|ALL|THRU|ZERO
Example: To query calibration step 'Open' status:
[:SENSe]:CORRection:COLLect:THRU
Cmd Description: Perform 'Thru' step of calibration.
Example: To measure Thru:
[:SENSe]:CORRection:COLLect:TYPE
RFP1|2PFP|RRP1|TRFP|2PES|TRES
[:SENSe]:CORRection:COLLect:TYPE?
Cmd Description: Command sets calibration collection type.
Query Description: Query returns calibration collection type.
Param Range: RFP1|2PFP|RRP1|TRFP|2PES|TRES, Classic Cable & Antenna
Default Value: RFP1
Example: To set collection type to RFP1:
[:SENSe]:CORRection:COLLect:ZERO
Cmd Description: Perform 'Zero Sensor' step of calibration.
Example: To zero sensor:
[:SENSe]:CORRection:INSTacal:CALibrate
Cmd Description: Performs InstaCal.
Example: To perform a InstaCal:
S331L PM
Otherwise, it returns 0. Query without any parameter returns 0 if no cal
started, 1 if cal already started, 2 if cal has been aborted, 3 if cal is
calculating, or 4 if cal has been completed.
:CORR:COLL:STAT? OPEN
:CORR:COLL:THRU
Analyzer Mode only supports RFP1 and TRES
:CORR:COLL:TYPE RFP1
:CORR:COLL:ZERO
:CORR:INST:CAL
PN: 10580-00322 Rev. F
4-25